| E1 | 100 g — 1 kg |
| E2 | 100 g — 1 kg |
| F1 | 100 g — 1 kg |
| F2 | 100 g — 1 kg |
| Maximum capacity [Max] | 1.02 kg |
| Readability [d] | 0.005 mg |
| Standard repeatability [5% Max] | 8 µg |
| Standard repeatability [Max] | 15 µg |
| Permissible repeatability | 25 µg |
| Eccentricity (tested load) | 0 mg |
| Electric compensation range | -10 g — +20 g |
| Stabilization time | 30 s |
| Adjustment | external |
| Display | 10″ touchscreen |
| Magazine positions | 4 |
| Weighing pan dimensions | ø50 mm |
| Weighing device dimensions | 385×215×600 mm |
| Controlling device dimensions | 206×140×70 mm |
| Anti-draft chamber dimensions | 560×340×665 mm |
| Packaging dimensions | 860×800×560 mm |
| Chamber packaging size | 950×420×630 mm |
| Net weight | 47.8 kg |
| Gross weight | 67.8 kg |
| Supplementary weights internal | half automatic |
| Supplementary weights external | – |
| Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
| Power supply | 100-240 V AC 50/60 Hz |
| Operating temperature | +15 — +30°C |
| Operating temperature change rate | ±0.5°C/12h (±0.3°C/4h) |
| Relative humidity | 40% — 60% |
| Relative humidity change rate | ±5%/12h (3%/4h) |
Repeatability is expressed as a standard deviation determined for 6 ABBA cycles. Standard deviation is experimentally determined under ambient conditions for calibration of E1 class mass standards specified in OIML R111 (Table C.1.) document.



![Timbangan-Supmeter-BST106-M10[FL]](https://mitrahitech.com/wp-content/uploads/2019/08/Timbangan-Supmeter-BST106-M10FL-247x247.jpg)
![Timbangan-Supmeter-BST106-H16[AD]](https://mitrahitech.com/wp-content/uploads/2019/08/Timbangan-Supmeter-BST106-H16AD-247x247.jpg)

![Timbangan-Supmeter-BST106--B60R[F]](https://mitrahitech.com/wp-content/uploads/2019/08/Timbangan-Supmeter-BST106-B60RF-247x247.jpg)
![Timbangan-Supmeter-BST106-M10[AL]](https://mitrahitech.com/wp-content/uploads/2019/08/Timbangan-Supmeter-BST106-M10AL-247x247.jpg)

Ulasan
Belum ada ulasan.