FEATURES
ELECTRONIC LEVEL INDICATOR
Functions:
- ALARM function
- graphic level indicator
- programmable acceptable tilts
DATA EXCHANGE THROUGH USB STORAGE DEVICES
- update balance software
- export weighing data
- export/import databases
- export/import balance settings
- exchange data between balances
INFRARED PROXIMITY SENSORS
Optional functions:
- PRINT function
- TARE function
- sensor’s sensitivity adjustment
PORTABILITY
Application:
Elimination of extra sources of vibration
Using wireless terminal facilitates elimination of additional vibrations resulting from daily operation on a balance. Owing to that, balances with highest resolutions might be used without the need of dedicated weighing workstations.
Convenience and comfort of weighing in fume cupboards and laminar flow cabins
Wireless communication between a terminal and a weighing module makes balance operation more comfortable and safer.
Balance isolation form influence of environmental conditions
Wireless communication allows to either place the weighing module inside a draft shield or to isolate it from the operator. This allows to weight toxic substances without putting operator’s health or life at risk.
Enhanced pipetting ergonomics
Pipettes calibration process requires both, measurements precision and fast operation. Radwag wireless communication creates a new dimension of comfort for your daily enjoyment.
INTERFACES
- Ethernet
- 2×USB
- 2×RS 232
- 4 in / 4 out
- WiFi 2,4GHz b,g,n
FUNCTIONS
AUTOTEST
Diagnostic function aiming at metrological parameters determination (repeatability), the parameters are determined for the actual conditions of use. When speaking of repeatability it may be also used for weighing time optimization. Autotest is operated in an automatic mode thus operator’s time is saved.
DOSING
Weighing process for which reference mass has been determined together with tolerance for its determination. Dosing tolerance is given in [%] and it is calculated in relation to the reference value thus being a permissible deviation of this process. This solution is used for weighing powders, liquids and loose materials. Dosing function performance is often supported with bargraph – load indicator. For industrial scales it is possible to use a control systems of dosing process.
PERCENT WEIGHING
Percent Weighing function is used for comparision of measured products with mass standard. Mass of a mass standard may be a numeric value taken from a database or it may be determined through a measurement process. Each measured product is compared to mass standard, mass of which is presumed as a model 100% ideal mass. For products weighing less than the mass standard, obtained results are lower than 100%, for products weighing more, the obtained results are greatly exceeded.
PARTS COUNTING
Function using mass measurement for determination of measured items quantity. Mass of a single item is required for this process. It may be either estimated through weighment or taken from a database. For items counting the following algorithm is used: all items mass / single item mass = quantity. Function operation is supperted by a mechanism of Automatic Correction of Accuracy. This allows to update single item mass in course of the process. To a certain extend Automatic Correction of Accuracy eliminates error which may be a result of different mass values of seemingly alike single elements. For industry solutions items counting may be simultaneously carried out with checkweighing and dosing thus industry solutions feature audio signalling base informing that specified number of items has been weighed. It is possible to apply weighing systems using few platfroms of different MAX capacities and different accuracies.
FORMULATION
Function supporting the mixture making process, wherein the mixture contains various components. Formulation function usually uses the balance/scale database of components. Formulation serves for monitored checkweighing of every single component with a given tolerance. It is enriched with a set of individual settings.
STATISTICS
Statistiscs function registers and analyses performed measurements. This supplies the user withthe following information: Max and Min standard deviation, average value, variance, range et.
CHECKWEIGHING
Checkweighing function is used for checking whether the measured sample mass is within the predefined threshold values, Low [LO] and Hgh [HI]. The thresholds are given in [g] and [kg] units. Current state of a sample being measured is signalled by means of pictorgams located on a display for laboratory balances, for industrial scales Stackligt System is used. This visual +/- inspection is in operation during segregation, control or packing process of products for which mass has bees determined with a specifaied tolerance, eg. 12860 g 961
IR SENSORS
Programmable function supporting the weighing process through control of the following options: sliding weighing chamber doors, printout, zeroing, tarring etc. Especially appreciated wherever preventing the balance from soiling is important.
UNDER-PAN WEIGHING
The solution enables weighing with use of a stand, which is advisable for loads of non-standard shapes and dimensions and those that generate magnetic field.
GLP PROCEDURES
Diagnostic function allowing to objectively document performed measurements. GLP procedures may be either presented in a short report form or extended one.
ANIMAL WEIGHING
Process of mass determination for a product which may unwillingly reposition within the weighing pan. Mass determination in such cases requires much longer period of time when compared to typical weighing process. It is the user who defines period of time needed for control of measured product mass. The user can thus optimize the function depending on the measured product characteristics.
AIR DENSITY CORRECTION
Function performing correction of mass measurement indication, wherein the air density is taken into account. It is used in balances with reading unit < 0,01 mg.
DENSITY DETERMINATION
Liquid or solid body density function uses the Archimedes principle and it requires a special kit to be applied. The kit when used, replaces a weighing pan. For solids mass of the sample must be determined first in the air, next in the liquid of specified density. For liquids the so called sinker, density of which has been determined, must be measured first in the air, next in the tested liquid. The density determination process is disturbed by air bubbles which may stick either to the measured sample or the sinker. In case of laboratory equipment very light solids may be measured. For industry, solid bodies density may be determined by means of under-pan weighing. Tested load of significant size may be measured this way
DIFFERENTIAL WEIGHING
Differential weighing informs about mass variation a particular sample, wherein the said variation is a result of manufacturing processes.
Differential weighing is often applied in product control process, for which reference product parameters have been specified in order to be used as a source of comparison for the whole series.
Insuchcasestandard deviation is the information which tells about quality and completeness of a final product. For systems based on a feed-back technique such information may be used by control units.
REPLACEABLE UNIT
Replaceable units.
STATISTICAL QUALITY CONTROL
Statistical control for determination of basic statistical criteria: maximum, minimum, standard deviation, mean values for each batch, etc.
WI-FI®
Wi-Fi® offers possibility to send data from balance to a mobile iOS or Android device, via special data management apps. Wireless communication between the indicator and the weighing instrument enables operating the balance that is placed inside a laminar flow hood or a fume cupboard.
DATASHEET
Maximum capacity [Max] | 310 g |
Minimum load | 10 mg |
Readability [d] | 0.1 mg |
Tare range | -310 g |
Minimum weight (USP) | 160 mg |
Minimum weight (U=1%,k=2) | 16 mg |
Repeatability (Max) | 0.15 mg |
Repeatability (5% Max) | 0.08 mg |
Linearity | ±0.3 mg |
Stabilization time | 3.5 s |
Adjustment | internal (automatic) |
OIML Class | I |
Display | 5.7” resistive colour touchscreen |
Communication interface | 2×RS232, 2×USB-A, Ethernet, 4 IN / 4 OUT (digital), Wi-Fi® |
Power supply | 100 ÷ 240 V AC 50 / 60 Hz |
Power consumption max. | 7 W |
Operating temperature | +10 ÷ +40 °C |
Weighing pan dimensions | ø100 mm |
Packaging dimensions | 600×400×540 mm |
Net weight | 6.2 kg |
Gross weight | 11 kg |
ACCESSORIES
- Granite Antivibration tables
- Antivibration tables for laboratory balances
- Professional weighing table
- Audit Trial Reader
- E2R System
- Label Editor R02
- 3Y Database Editor
- Parameter Editor
- R.Barcode
- R-LAB
- RAD-KEY
- RADWAG Connect
- RADWAG Development Studio
- RADWAG Remote Desktop
- labVIEW Driver
- Under-Pan Weighing Rack
- Density determination kit
- Receipt Printer
- Label Printers
- Antistatic Ionizer
- THBR 2.0 System – Ambient Conditions Monitoring
- Cigarette lighter receptacle power
- RS 232 cables (scale-EPSON Printer)
- USB Hubs
- Moxa Uport 1110 Converter
- RS 232, RS 485 cables
- Protective cover for balances
- Automatic Feeders
- Barcode Scanners
- Holder for laboratory flasks
- Holder for test tubes and filters
- Displays
- Power Adapters
ADDITIONAL MODULES
21 CFR Part 11 Compliance Module
The 21 CFR Part 11 Compliance Module is a solution especially useful for those users who are obliged to conform to provisions of 21 CFR Part 11 and EU GMP Part 4, Annex 11 regulations. These two documents determine the rules for using electronic records and signatures.
Data protection is guaranteed by:
- advanced password settings,
- control of logging correctness, various permission levels,
- data management access,
- backup,
- electronic signature,
- time-defined automatic logout,
- modification record to Audit Trail file.
The module features “Audit Trail” which registers each data modification (date, start value, modified value, operator responsible for the modification). This enables tracking all activities and modifying them in the future if required.
The Audit Trail Reader freeware allows users to acquire data from the weighing instrument and export it as files to pdf, xls, csv, html formats.
Ulasan